최소 단어 이상 선택하여야 합니다.
최대 10 단어까지만 선택 가능합니다.
다음과 같은 기능을 한번의 로그인으로 사용 할 수 있습니다.
NTIS 바로가기Progress in superconductivity, v.13 no.1, 2011년, pp.1 - 6
이용호 (Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science) , 김진목 (Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science) , 유권규 (Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science) , 김기웅 (Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science) , 권혁찬 (Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science)
Measurement of magnetic signals generated from electric activity of myocardium provides useful information for the functional diagnosis of heart diseases. Key technical component of the magnetocardiography (MCG) technology is SQUID. To measure MCG signals with high signal-to-noise ratio, sensitive S...
* AI 자동 식별 결과로 적합하지 않은 문장이 있을 수 있으니, 이용에 유의하시기 바랍니다.
핵심어 | 질문 | 논문에서 추출한 답변 |
---|---|---|
SQUID는 무엇인가? | SQUID는 기본적으로 자속신호를 전압으로 변환시키는 자속-전압변환 소자이다. 현재 대부분 사용되는 DC-SQUID는 자속-전압 변환계수가 약 100 μV/ Φ0 내외이다. | |
SQUID 자기센서의 감도가 우수해야 하는 이유는 무엇인가? | 심자도 파형의 검출을 위해 단순히 P, QRS, T-파의 피크를 검출하는 데에는 고감도 자기센서가 필요하지 않지만, 파형의 미세한 차이를 정밀하게 분석하든가 작은 부위에서 발생되는 미세신호를 측정하기 위해서는 SQUID 자기센서의 감도가 우수해야 한다. 또한 다채널 센서 장치에 사용되기 위해서는 SQUID의 제작수율 및 센서 파라메터 조절의 재현성, 장기간 동작에 대한 물리적인 신뢰성 등이 양호해야 한다. | |
additional positive feedback(APF) 방식은 어떤 방식인가? | 자속-전압변환계수를 증가시키는 방법 중의 대표적으로 많이 사용되고 있는 방법은 DC SQUID에 additional positive feedback(APF)을 추가하는 것이다. APF 방식은 DC SQUID의 자속-전압변환계수를 측정한 후 SQUID와 병렬로 저항과 코일로 구성된 회로를 부착하여 바이어스 전류의 일부가 자기장으로 SQUID loop에 전달되도록 하여 자속-전압곡선을 비대칭적으로 하여 자속-전압변환계수를 증가시키는 방법이다. 이 방법은 SQUID 특성 평가 후 최적의 피드백 저항-코일을 계산하여 부착해야 하는 단점이 있다. |
R. Fenici, D. Brisinda and A. M. Meloni, "Clinical application of magnetocardiography", Expert Rev. Mol. Diagn. 5, 291-313 (2005).
S. Yamada S and I. Yamaguchi, "Magnetocardiograms in Clinical Medicine: Unique Information on Cardiac Ischemia, Arrhythmias, and Fetal Diagnosis", Internal Medicine, 44, 1-19 (2005).
J. W. Park and F. Jung, "Qualitative and quantitative description of myocardial ischemia by means of magnetocardiography", Biomed. Technik. 49, 267-273 (2004).
D. Cohen, E. A. Edelsack and J. E. Zimmerman, "Magnetocardiograms taken inside a shielded room with superconducting point-contact magnetometer", Appl. Phys. Lett., 19, 278 (1970).
K. Sternikel and A. I. Braginski, "Biomagnetism using SQUIDs: status and perspectives", Supercond. Sci. Technol., 14, 160-171 (2006).
I. Tavarozzi, S. Comani, C. Dell Gratta, G. L. Romani, S. Di Luzio, D. Brisinda, S. Gallina, M. Zimarino, R. Fenici, R. De Caterina, "Magnetocardiography: current status and perspectives. Part I: Physical principles and instrumentation", Ital. Heart J., 3, 75-85 (2002).
H. Nowak, "Magnetism in Medicine", ed W. Andra and H. Nowak, Berlin: Wiley, 85-135 (1998).
M.B. Ketchen, "Integrated thin-film dc SQUID sensors", IEEE Trans. Magn., MAG-23, 1650-1657 (1987).
Y. H. Lee, K. K. Yu, J. M. Kim, H. Kwon, K. Kim and Y. K. Park, "64-channel second-order axial gradiometer system based on DROS for magnetocardiogram in a thin shielded room", Physica C, 468, 1942-1945 (2008).
D. Drung, "SQUID Sensors: Fundamentals, Fabricati on and Applications", ed H. Weinstock, Dordrecht: K luwer, pp 63-116 (1996).
Y. H. Lee, H. C. Kwon, J. M. Kim, Y. K. Park and J. C. Park, "Double relaxation oscillation SQUID with reference junction for biomagnetic multi-channel applications", Appl. Supercond., 5, 413-418 (1998).
Y. H. Lee, J. M. Kim, H. Kwon, K. K. Yu, I. S. Kim a nd Y. K. Park, "64-channel magnetocardiogram syste m based on double relaxation oscillation SQUID plan ar gradiometers", Supercond. Sci. Technol. 19 S284-S288 (2006).
Y. H. Lee, K. K. Yu, J. M. Kim, H. Kwon and K. Kim, "A 64-channel MCG system having divided gradiometer array inside a low boil-off dewar", Supercond. Sci. Technol., 22, 1-7 (2009).
*원문 PDF 파일 및 링크정보가 존재하지 않을 경우 KISTI DDS 시스템에서 제공하는 원문복사서비스를 사용할 수 있습니다.