호흡모의움직임장치를 이용한 정상산소 폴리메타크릴산 겔 선량계의 선량특성 Preliminary Investigations of the Dosimetric Properties of a Normoxic Polymethacrylic Acid Gel Dosimeter Using a Respiration-Motion Simulator원문보기
Dose distribution throughout the clinical organ range of motion was analyzed using a respiratory-motion simulator that was equipped with a polymer gel dosimeter and EBT2 film. The normoxic polymer gel dosimeter was synthesized from gelatin, MAA, HQ, THPC and HPLC. The gel dosimeter and EBT2 film wer...
Dose distribution throughout the clinical organ range of motion was analyzed using a respiratory-motion simulator that was equipped with a polymer gel dosimeter and EBT2 film. The normoxic polymer gel dosimeter was synthesized from gelatin, MAA, HQ, THPC and HPLC. The gel dosimeter and EBT2 film were irradiated with Co-60 gamma rays that were moved along the x-axis and y-axis in ${\pm}1.5cm$ steps at five-second intervals. The field size was $5{\times}5cm^2$. The SSD was 80 cm and set to 10 Gy at a depth of 2 cm. The PDD at a depth of 50 mm was 75.2% in the ion chamber, 82.3% in the static state and 86.1% in the dynamic state in the gel dosimeter. The penumbra for the dynamic state target, which was measured using the gel dosimeter, averaged 10.89 mm, this is a 40.5% increase over the penumbra of the static state target of 7.74 mm. In addition, when measuring with gel dosimetry, the value for the penumbra is 36.6% smaller in the static state and 29.4% smaller in the dynamic state compared to measuring with film. The aim of this study was to investigate the dosimetric properties of a normoxic polymethacrylic acid gel dosimeter in static and dynamic states and to evaluate the potentiality as a relative dosimeter for dynamic therapeutic radiation.
Dose distribution throughout the clinical organ range of motion was analyzed using a respiratory-motion simulator that was equipped with a polymer gel dosimeter and EBT2 film. The normoxic polymer gel dosimeter was synthesized from gelatin, MAA, HQ, THPC and HPLC. The gel dosimeter and EBT2 film were irradiated with Co-60 gamma rays that were moved along the x-axis and y-axis in ${\pm}1.5cm$ steps at five-second intervals. The field size was $5{\times}5cm^2$. The SSD was 80 cm and set to 10 Gy at a depth of 2 cm. The PDD at a depth of 50 mm was 75.2% in the ion chamber, 82.3% in the static state and 86.1% in the dynamic state in the gel dosimeter. The penumbra for the dynamic state target, which was measured using the gel dosimeter, averaged 10.89 mm, this is a 40.5% increase over the penumbra of the static state target of 7.74 mm. In addition, when measuring with gel dosimetry, the value for the penumbra is 36.6% smaller in the static state and 29.4% smaller in the dynamic state compared to measuring with film. The aim of this study was to investigate the dosimetric properties of a normoxic polymethacrylic acid gel dosimeter in static and dynamic states and to evaluate the potentiality as a relative dosimeter for dynamic therapeutic radiation.
* AI 자동 식별 결과로 적합하지 않은 문장이 있을 수 있으니, 이용에 유의하시기 바랍니다.
가설 설정
12) The penumbra area is the edge of the radiation beam, over which the dose rate changes rapidly as a function of distance from the beam axis.
제안 방법
The gel dosimeter and film were installed in a respiratory-motion simulator, and the dose distributions in the static and dynamic states were measured. After irradiation with Co-60 gamma rays, an image was obtained using MRI, and the percent depth dose and penumbra effect were analyzed.
2. Experimental setup of the Co-60 teletherapy unit with a gel dosimeter and a respiratory-motion simulator for static and dynamic dosimetry.
The preliminary investigations of a polymethacrylic acid gel for PDD, flatness, symmetry, and penumbra were performed in order to find its applicability to therapeutic dynamic targets.
대상 데이터
The polymer gel dosimeter presented the PDD with respect to the central axis for the image obtained using MRI. The PDD was presented in a range between 10 mm and 60 mm.
참고문헌 (15)
Kubo HD, Len PM, Minohara S, et al: Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center. Med Phys 27:346-353 (2000)
Kim, YL, Chung JB, Chung WK, et al: An Effect of Time Gating Threshold (TGT) on the Delivered Dose at Internal Organ with Movement due to Respiration. Korean J Med Phys 16:89-96 (2005)
Park SW, Jung HJ, Kim KB, et al: Development and Evaluation of a Target-tracking Radiation-therapy system Using a Multileaf Collimator (MLC) Synchronized with Moving Organs. J Korean Phys Soc 55:694-701 (2009)
Kanagaki B, Read PW, Molloy JA, Larner M, Sheng K: A motion phantom study on helical tomotherapy: the dosimetric impacts of delivery technique and motion. Phys Med Biol 52: 243-255 (2007)
Ceberg S, Karlsson A, Gustavsson H, et al: Verification of dynamic radiotherapy:the potential for 3D dosimetry under respiratory-like motion using polymer gel. Phys Med Biol 53: N387-N396 (2008)
Kim MS, Ha S, Lee DH, et al: Distribution of the dose profile in a three-dimensional moving phantom to simulated tumor motion during image-guided radiosurgery. Radiation Oncology Journal 25(4):268-277 (2007)
Suh Y, Yi B, Ahn S, Kim J, Lee S, Shin S, Choi E: Aperture maneuver with compelled breath for moving tumors: A feasibility study with a moving phantom. Med Phys 31(4): 760-766 (2004)
Cho YR, Park HW, Kim AR, et al: Fabrication of a normoxic polymer gel dosimeter and its dose distribution characteristics. J Korean Phy Soc 59(1): 169-175 (2011)
Fuss M, Sturtewagen E, De Wagter C, Georg D: Dosimetric characterization of Gafchromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 52(14):4211-4225 (2007)
Nath R, Biggs PJ, Bova FJ, et al: AAPM code of practice for radiotherapy accelerators: report of AAPM radiation Therapy Task Group No. 45. Med Phys 21:1093-1121 (1994)
Cho KH, Cho SJ, Lee S, et al: Dose responses in a normoxic polymethacrylic acid gel dosimeter using optimal CT scanning parameters. Nuclear Instruments and Methods in Physics Research A 675:112-117 (2012)
※ AI-Helper는 부적절한 답변을 할 수 있습니다.