권순무 (Daegu Health College Dept.of Radiologic Technology) , 박정규 (Daegu Health College Dept.of Radiologic Technology) , 김부순 (Daegu Health College Dept.of Radiologic Technology)
The angular dependence of active dosimeters, EPD, is analysed and compared with that of passive dosimeters, OSLD, after evaluating their relative response and uncertainty of measurement, where it is known that the personal use of them has been increased recently. There appeared a minor variation for...
* AI 자동 식별 결과로 적합하지 않은 문장이 있을 수 있으니, 이용에 유의하시기 바랍니다.
핵심어 | 질문 | 논문에서 추출한 답변 |
---|---|---|
TLD의 장점은 무엇인가? | 방사선을 이용한 의료영상분야에서 작업종사자들의 방사선 피폭을 측정하기 위한 법정선량계로 film badge, thermo luminescent dosimeter(TLD), radiophoto luminescent glass dosimeter (RPLD) 등이 사용되고 최근에는 optical stimulated luminescence dosimeter(OSLD)의 사용이 증가하고 있다[1]. 이러한 수동형 선량계 중 TLD는 넓은 선량측정범위(수 μSv~수천 Sv), 우수한 감도, 장기간의 집적선량측정, 낮은 온도 변동성(±5% 이내), 높은 정확도와 재현성(±3% 이내)으로 일반적으로 사용되는 개인피폭선량계이다[2, 3]. OSLD는 방사선에 조사된 알루미늄산화물(Al2O3:C)에 특정 파장의 광자극으로 형광을 발생시키고 이 형광을 광전자증배관으로 수집한다. | |
전자식개인선량계의 장점은 무엇인가? | 이러한 수동형 선량계의 단점을 극복하고 효율적인 방사선 피폭선량 측정이 가능한 능동형 선량계인 전자식개인선량계 (electronic personal dosimeter, 이하EPD)의 보급과 사용이 증가하고 있다. EPD는 즉각적인 선량 및 선량률 표시기능, 일정선량 이상에서 알람을 울리는 경보기능, 피폭이력 제공 및 효율적인 선량 자료 관리 등의 장점이 있다. 또한 수동형 선량계보다 3배 이상 더 낮은 저선량을 측정할 수 있어 보조선량계 역할을 하고있다[7, 8]. 이런 장점으로 프랑스에서는 법정 선량계로서 사용되고 있으며 영국에서는 사용이 증가하고 있다[9]. | |
수동형 선량계의 단점은 무엇인가? | 또한 정확성과 재현성이 우수하며, 방사선을 사용하는 다양한 분야에 임상목적으로 적용이 가능하다고 보고되고 있다[6]. 하지만 수동형 선량계는 즉각적인 판독이 어렵고, 재사용을 위한 열처리(annealing) 과정이 필요하며 피폭이력 확인이 불가능하다는 단점이 있다. 이러한 수동형 선량계의 단점을 극복하고 효율적인 방사선 피폭선량 측정이 가능한 능동형 선량계인 전자식개인선량계 (electronic personal dosimeter, 이하EPD)의 보급과 사용이 증가하고 있다. |
W.H. Lee, S.C Kim, and S.M Ahn. "Comparison on the Dosimetry of TLD and OSLD Used in Nuclear Medicine." The Journal of the Korea Contents Association Vol12. No2. pp329-334. 2012
Essers M, and BJ Milnheer. "In vivo dosimetry during external photon beam radiotheraphy", International Journal of Radiation Oncology, Biology, Physics, Vol43. No2. pp245-255. 1999
J.H.Choi, G.J. Kang, and S.G. Chang. "Comparison on the dosimetry of TLD and PLD by dose area product." The Journal of the Korea Contents Association Vol12. No3. pp244-250. 2012
Jursinic PA. Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med Phys. 2007;34(12):4594-4604
K. S. Chae, Y. K. Kim, S. I. Ho, K. S. Nho. "A Study on Thermoluminescent Dosimeter" kyungnam university. materials research Vol5. pp143-151. 1995
I. Mrcela, T. Bokulic, and J. Izewska,"Optically stimulated luminescence in vivo dosimetry for radio therapy: physical characterization and clinical measurements in 60Co beams," Physics in Medicine and Biology, Vol.56. No.18. 2011.
M Luzik-Bhadra and S Perle. "Electronic personal dosemeters will replace passive dosemeters in the near future". Radiat Prot Dosimetry. Vol123. No4. pp546-553. 2007
Bolognese-Milsztajn T, Ginjaume M, Luszik-Bhadra M, Vanhavere F, Wahl W and Weeks A. "Active personal dosimeters for individual monitoring and other New Developments". Radiat Prot Dosimetry. Vol12. No1. pp141-168. 2012
Zodiates T. "Operational experience with a legal electronic dosimetry system". 4th European ISOE, Lyon (France), 2004
JP McCaffrey, H Shen and B Downton. "Dose rate dependency of electronic personal dosemeters measuring X- and $\gamma$ -ray radiation. Radiat. Prot. Dosim". Radiat Prot Dosimetry. Vol131. No2. pp229-235. 2008
J.S. Seo etc. "Electronic personal dosimeter for gamma-ray measurement angular-dependent experiments", The korean association for radiation protection, pp 186-187. 2009
korean standards association, "IEC 61267 Medical diagnostic X-ray equipment-Radiation conditions for use in the determination of characteristics".
Perks, Christopher A., Cliff Yahnke, and Marc Million. "Medical dosimetry using Optically Stimulated Luminescence dots and microStar readers." 2008..
http://www.thermoscientific.com/en/product/epdelectronic-personal-dosimeters.html
IEC 61526 "Radiation protection instrumentation - direct reading personal dose equivalent meters and monitors".
Nuclear safety and swcurity commission http://www.nssc.go.kr/nssc/information/law2.jsp?modeview&article_no3012&pager.offset100&board_no9
M. S. Akselrod, Btter-Jensen L, and S.W. McKeever "Optically stimulated luminescence and its use in medical dosimetry", Radiat Measurements. Vol. 41, pp.78-99, 2006.
"Radiation detection & measurement", 2012, ISBN-13 978-89-5616-498-4, Chungu publishing co.
J. R. Kerns, S. F.Kry, Sahoo, N. Followill, D. S., &Ibbott, G. S. "Angular dependence of the nanoDot OSL dosimeter". Medical physics, Vol38. No7. pp3955-3962. 2011