Overcoming the problem of residual microbial contamination in dental suction units left by conventional disinfection using novel single component suction handpieces in combination with automated flood disinfection원문보기
Objectives: Decontaminating dental chair unit (DCU) suction systems in a convenient, safe and effective manner is problematic. This study aimed to identify and quantify the extent of the problems using 25 DCUs, methodically eliminate these problems and develop an efficient approach for reliable, eff...
Objectives: Decontaminating dental chair unit (DCU) suction systems in a convenient, safe and effective manner is problematic. This study aimed to identify and quantify the extent of the problems using 25 DCUs, methodically eliminate these problems and develop an efficient approach for reliable, effective, automated disinfection. Methods: DCU suction system residual contamination by environmental and human-derived bacteria was evaluated by microbiological culture following standard aspiration disinfection with a quaternary ammonium disinfectant or alternatively, a novel flooding approach to disinfection. Disinfection of multicomponent suction handpieces, assembled and disassembled, was also studied. A prototype manual and a novel automated Suction Tube Cleaning System (STCS) were developed and tested, as were novel single component suction handpieces. Results: Standard aspiration disinfection consistently failed to decontaminate DCU suction systems effectively. Semi-confluent bacterial growth (101-500 colony forming units (CFU) per culture plate) was recovered from up to 60% of suction filter housings and from up to 19% of high and 37% of low volume suction hoses. Manual and automated flood disinfection of DCU suction systems reduced this dramatically (ranges for filter cage and high and low volume hoses of 0-22, 0-16 and 0-14CFU/plate, respectively) (P<0.0001). Multicomponent suction handpieces could not be adequately disinfected without prior removal and disassembly. Novel single component handpieces, allowed their effective disinfection in situ using the STCS, which virtually eliminated contamination from the entire suction system. Conclusion: Flood disinfection of DCU suction systems and single component handpieces radically improves disinfection efficacy and considerably reduces potential cross-infection and cross-contamination risks. Clinical significance: DCU suction systems become heavily contaminated during use. Conventional disinfection does not adequately control this. Furthermore, multicomponent suction handpieces cannot be adequately disinfected without disassembly, which is costly in time, staff and resources. The automated STCS DCU suction disinfection system used with single component handpieces provides an effective solution.
Objectives: Decontaminating dental chair unit (DCU) suction systems in a convenient, safe and effective manner is problematic. This study aimed to identify and quantify the extent of the problems using 25 DCUs, methodically eliminate these problems and develop an efficient approach for reliable, effective, automated disinfection. Methods: DCU suction system residual contamination by environmental and human-derived bacteria was evaluated by microbiological culture following standard aspiration disinfection with a quaternary ammonium disinfectant or alternatively, a novel flooding approach to disinfection. Disinfection of multicomponent suction handpieces, assembled and disassembled, was also studied. A prototype manual and a novel automated Suction Tube Cleaning System (STCS) were developed and tested, as were novel single component suction handpieces. Results: Standard aspiration disinfection consistently failed to decontaminate DCU suction systems effectively. Semi-confluent bacterial growth (101-500 colony forming units (CFU) per culture plate) was recovered from up to 60% of suction filter housings and from up to 19% of high and 37% of low volume suction hoses. Manual and automated flood disinfection of DCU suction systems reduced this dramatically (ranges for filter cage and high and low volume hoses of 0-22, 0-16 and 0-14CFU/plate, respectively) (P<0.0001). Multicomponent suction handpieces could not be adequately disinfected without prior removal and disassembly. Novel single component handpieces, allowed their effective disinfection in situ using the STCS, which virtually eliminated contamination from the entire suction system. Conclusion: Flood disinfection of DCU suction systems and single component handpieces radically improves disinfection efficacy and considerably reduces potential cross-infection and cross-contamination risks. Clinical significance: DCU suction systems become heavily contaminated during use. Conventional disinfection does not adequately control this. Furthermore, multicomponent suction handpieces cannot be adequately disinfected without disassembly, which is costly in time, staff and resources. The automated STCS DCU suction disinfection system used with single component handpieces provides an effective solution.
Am. J. Crit. Care Sole 11 141 2002 10.4037/ajcc2002.11.2.141 Bacterial growth in secretions and on suctioning equipment of orally intubated patients: a pilot study
Gen. Dent. Merchant 38 189 1990 Evacuation system lines and solid waste filter traps: associated flora and infection control
J. Hosp. Infect. O’Donnell 59 348 2005 10.1016/j.jhin.2004.10.001 Bacterial contamination of dental chair units in a modern dental hospital caused by leakage from suction system hoses containing extensive biofilm
J. Am. Dent. Assoc. Watson 124 77 1993 10.14219/jada.archive.1993.0080 Possibility of cross-contamination between dental patients by means of the saliva ejector
J. Hosp. Infect. Jensen 36 117 1997 10.1016/S0195-6701(97)90117-1 Epidemiology of Pseudomonas aeruginosa in cystic fibrosis and the possible role of contamination by dental equipment
J. Dent. O’Donnell 4 648 2006 10.1016/j.jdent.2005.12.006 A novel automated waterline cleaning system that facilitates effective and consistent control of microbial biofilm contamination of dental chair unit waterlines: a one-year study
J. Dent. O’Donnell 37 748 2009 10.1016/j.jdent.2009.06.001 A centralised, automated dental hospital water quality and biofilm management system using neutral Ecasol™ maintains dental unit waterline output at better than potable quality: a 2-year longitudinal study
J. Hosp. Infect. Tuttlebee 52 192 2002 10.1053/jhin.2002.1282 Effective control of dental chair unit waterline biofilm and marked reduction of bacterial contamination of output water using two peroxide-based disinfectants
J. Med. Microbiol. Ohara-Nemoto 57 95 2008 10.1099/jmm.0.47561-0 Occurrence of staphylococci in the oral cavities of healthy adults and nasal-oral trafficking of the bacteria
Mol. Microbiol. Carroll 16 877 1995 10.1111/j.1365-2958.1995.tb02315.x Novel organization of the site-specific integration and excision recombination functions of the Staphylococcus aureus serotype F virulence-converting phages phi 13 and phi 42
J. Med. Microbiol. Humphreys 28 163 1989 10.1099/00222615-28-3-163 Enterotoxin production by Staphylococcus aureus isolates from cases of septicaemia and from healthy carriers
PLoS One Monecke 6 e17936 2011 10.1371/journal.pone.0017936 A field guide to pandemic, epidemic and sporadic clones of methicillin-resistant Staphylococcus aureus
※ AI-Helper는 부적절한 답변을 할 수 있습니다.