NTIS 바로가기방사선기술과학 = Journal of radiological science and technology, v.46 no.3, 2023년, pp.219 - 229
목성윤 (동의대학교 방사선학과) , 강영록 (동남권원자력의학원) , 김효진 (동남권원자력의학원) , 계용욱 (동남권원자력의학원) , 안현 (동의대학교 방사선학과)
In this study, basic characteristics such as reproducibility, linearity, and directionality of RPL glass dosimeters were evaluated to improve the reliability of dose evaluation through RPL glass dosimeters, and uncertainty elements such as sensitivity by glass element and magazine slot sensitivity w...
Rah JE, Hong JY, Suh TS, Shin DO, Kim HS, Lim?CI, et al. Study on dosimetric properties of radiophotoluminescent?glass rod detector. Journal of the?Korean Association for Radiation Protection. 2006;31(4):181-6.
Rah JE, Suh WS, Shin DO, Kim HS, Suh TS.?Determination of output factors for the gamma knife?using a radiophotoluminescent glass rod detector.?Korean Journal of Medical Physics. 2007;18(1):13-9.
Mizuno H, Fukumura A, Fukahori M, Sakata S,?Yamashita W, Takase N, et al. Application of a radiophotoluminescent?glass dosimeter to nonreference?condition dosimetry in the postal dose audit system.?Medical Physics. 2014;41(11):112104-1-6.
Sergieva KM. Dosimetry audit in modern radiotherapy.?IntechOpen; 2021:69-87.
Kato M, Chida K, Moritake T, Sato T, Oosaka H,?Toyoshima H, et al. Direct dose measurement on?patient during percutaneous coronary intervention?procedures using radiophotoluminescence glass?dosimeters. Radiation Protection Dosimetry. 2017;175(1):31-7.
ISO 22127:2019(E). Dosimetry with radiophotoluminescent?glass dosimeters for dosimetry audit in MV X-ray?radiotherapy. IAEA; 2019.
Huang DYC, Hsu SM. Radio-photoluminescence?glass dosimeter (RPLGD). IntechOpen; 2011.
Glass Dosimeter GD-300 Series Instruction Manual.?Shizuoka: AGC TECHNO GLASS Co., Ltd.; 2012.
Glass Dosimetry Reader FGD-1000 Instruction Manual?Ver.2. Shizuoka: AGC TECHNO GLASS Co., Ltd.; 2012.
Musolino SV. Absorbed dose determination in external?beam radiotherapy: An international code?of practice for dosimetry based on standards of absorbed?dose to water; Technical reports Series No. 398. IAEA; 2000.
ISO 4037-1. X and gamma reference radiation for?calibrating dosemeters and doserate meters and for?determining their response as a function of photon energy-Part 1 Radiation characteristics and production methods. IAEA; 2019.
ISO 4037-2. X and gamma reference radiation for?calibrating dosemeters and doserate meters and for?determining their response as a function of photon?energy-Part 2 Dosimetry for radiation protection?over the energy ranges 8 keV to 1.3 MeV and 4 MeV to 9 MeV. IAEA; 1997.
Korean Medical Physics Society Quality Control?Committee. Standard measurement method of absorbed?dose in water for high-energy photon and?electron beams. Korean Medical Physics Society?Quality Control Committee Report. KSMP; 2010.
KRISS/SP-2010-105. Guidelines for expression of?uncertainty in measurement. Korea Research?Institute of Standards and Science; 2010.
Moon YM, Rhee DJ, Kim JK, Kang YR, Lee MW,?Lim H, et al. Reference dosimetry and calibration of?glass dosimeters for Cs-137 Gamma-rays. Progress?in Medical Physics. 2013;24(3):140-4.
Mizuno H, Kanai T, Kusano Y, Ko S, Ono M,?Fukumura A, et al. Feasibility study of glass dosimeter?postal dosimetry audit of high-energy radiotherapy?photon beams. Radiotherapy and Oncology. 2008;86(2):258-63.
Korea Radioisotope Association. Basic characteristics?of glass dosimeters for personal dosimetry.?Radioisotope Journal. 1993;8(3):32-7.
Mobit PN, Nahum AE, Mayles P. A Monte Carlo?study of the quality dependence factors of common?TLD materials in photon and electron beams. Physics?in Medicine & Biology. 1998;43(8):2015-32.
Son GH, Jeong HJ, Sin SH, Lee HH, Lee SH, Kim MS,?et al. Feasibility study of the radiophotoluminescent?glass dosimeter for high-energy electron beams.?Korean Journal of Medical Physics. 2011;22(1):52-8.
*원문 PDF 파일 및 링크정보가 존재하지 않을 경우 KISTI DDS 시스템에서 제공하는 원문복사서비스를 사용할 수 있습니다.
출판사/학술단체 등이 한시적으로 특별한 프로모션 또는 일정기간 경과 후 접근을 허용하여, 출판사/학술단체 등의 사이트에서 이용 가능한 논문
※ AI-Helper는 부적절한 답변을 할 수 있습니다.