C-arm의 Tube 위치에 따른 거리 및 방향별 피폭선량 비교 A Comparision of the Radiation dose by Distance and the Direction according to a Tube Position of the C-arm Unit원문보기
In operation room, the use of the C-arm unit is increasing. So, the radiation dose of the person who work in operation room was even more increased than before. Thus, this study is shown the measurement of expose dose and the way for decrease of the radiation dose by using the C-arm unit. The experi...
In operation room, the use of the C-arm unit is increasing. So, the radiation dose of the person who work in operation room was even more increased than before. Thus, this study is shown the measurement of expose dose and the way for decrease of the radiation dose by using the C-arm unit. The experiment was performed with the C-arm unit and used a phantom which is similar to tissue of the human body and fluoro-glass dosimeter for dose measurement. The expose dose were measured by the tube position(over tube, under tube) of the C-arm unit, distance(50, 100$\sim$200cm), direction(I, II, III, IV), runtime(1min, 3min), wearing of the apron. The radiation dose was decreased twice and three times at under tube rather than over tube. The I direction was measured 20$\sim$30% more than the others. The biggest expose dose is 50cm from center on distance. The expose dose is decreased to far from center. In case of Wearing of the apron, the radiation dose was decreased 60$\sim$90% by the distance. But there weren't change of the radiation dose by C-arm tube position. In present, by increasing the usage of the C-arm unit, the radiation dose is inevitable. So, this study recommends us to use the under tube of the C-arm unit. Also, Wearing of the apron is required for minimum of the radiation exposure.
In operation room, the use of the C-arm unit is increasing. So, the radiation dose of the person who work in operation room was even more increased than before. Thus, this study is shown the measurement of expose dose and the way for decrease of the radiation dose by using the C-arm unit. The experiment was performed with the C-arm unit and used a phantom which is similar to tissue of the human body and fluoro-glass dosimeter for dose measurement. The expose dose were measured by the tube position(over tube, under tube) of the C-arm unit, distance(50, 100$\sim$200cm), direction(I, II, III, IV), runtime(1min, 3min), wearing of the apron. The radiation dose was decreased twice and three times at under tube rather than over tube. The I direction was measured 20$\sim$30% more than the others. The biggest expose dose is 50cm from center on distance. The expose dose is decreased to far from center. In case of Wearing of the apron, the radiation dose was decreased 60$\sim$90% by the distance. But there weren't change of the radiation dose by C-arm tube position. In present, by increasing the usage of the C-arm unit, the radiation dose is inevitable. So, this study recommends us to use the under tube of the C-arm unit. Also, Wearing of the apron is required for minimum of the radiation exposure.
* AI 자동 식별 결과로 적합하지 않은 문장이 있을 수 있으니, 이용에 유의하시기 바랍니다.
문제 정의
본 연구에서는 X선 tube의 위치에 따른 거리 및 방 향별 선량과 동일한 거리에서의 X선 방어용 apron 착 용 시 선량을 측정하여 수술팀의 피폭선량에 대해 알아 보고 더 나아가 피폭선량의 감소화 방안에 대한 연구를 진행하였다.
제안 방법
2) C-arm의 조사 중심으로부터 시계 방향으로 9시 방 향을 I, 12시 방향을 n, 3시 방향을 m, 6시 방향을 IV로 지정하고 각 방향의 50cm, 100cm, 200cm 거 리에서 방사선 감수성이 높은 갑상선을 피폭선량의 지표로 설정하여 한국 성인 남성 표준 신장을 기준으로 한 평균적 갑상선의 위치인 150cm 지점에 유 리선량계를 설치하였다(Fig. 1, Fig. 2)
4) 조사방법은 CINE mode를 이용하여 71 ~72kVp, 9.2~10mA로 over tube와 under tube에서 1분, 3 분간 조사하여 측정하고 위 방법과 동일하게 3회 실험하여 결과 값을 비교 분석하였다.
수술팀의 방사선 피폭선량을 알아보기 위하여 가상의 시술자(I), 마취과의사(H), 방사선사(m), 소독간호사 (N)의 근무 위치를 기준으로 하였다.
성능/효과
3). 50cm에서 3분 조사 시 over tube일 때의 거리별 평균 피폭선량은 0.24mGy로 측정되 었고 under tube일 때에는 Q059mGy로 over tube에서의 피폭선량이 under tube에서의 피폭선량 보다 4배 이상높게 나타나 동일한 결과를 나타내었다. 200cm에서의 피 폭선량은 1분 조사인 경우와 3분 경우인 때와 비교해서 별 차이가 없었다(Fig.
028mGy로 over tube 에서의 피폭선량이 under tube에서의 피폭선량 보다 3배 이상 높게 측정되었다. 거리가 증가할수록 측정된 선량 은 감소되었으며 200cm에서는 50cm의 10.1%로 선량이 급격히 감소하였다(Fig. 3). 50cm에서 3분 조사 시 over tube일 때의 거리별 평균 피폭선량은 0.
특히, 50cm에서 3배 이상 over tube 가 under tube위치에서 보다 피폭선량이 높게 나타났다. 방향별(I, D, HI, IV)로는 I방향이 다른 방향에 피해 피폭선량이 높게 나타났으며, X선 방어용 apron착용 시 는 tube 위치에 따른 피폭선량 차이는 크게 나타나지는 않았다(Table 3).
본 연구에서 C-am의 tube 위치에 따른 거리 및 방향 별 피폭선량을 측정한 결과, 100cm 이상의 거리에서부터 선량이 감소되어 200cm 거리에서는 tube 위치에 따른 선량이 큰 차이가 없었지만 50cm 거리에서는 피폭선량 이 3배 이상 차이가 나타났다. 대부분 수술실에서 Carm을 사용할 경우 시술자의 불편함으로 인해 under tube 방식보다는 over tube 방식을 선호하는 경우가 많 다.
C-arni의 위치에 따른 거리별 피폭선량에는 over tube 와 under tube 사이에 차이가 있었다. 특히 50cm에서 1 분 조사 시 over tube일 때 평균 피폭선량은 0.093mGy 로 측정 되었고 under tube일 때 0.028mGy로 over tube 에서의 피폭선량이 under tube에서의 피폭선량 보다 3배 이상 높게 측정되었다. 거리가 증가할수록 측정된 선량 은 감소되었으며 200cm에서는 50cm의 10.