C-arm을 이용한 EVAR(Endovascular Aneurysm Repair) 시술시 Lead Curtain 설치에 따른 Staff의 피폭선량 감소에 대한 연구 The investigation of the exposure dose reduction of the Staff according to the Lead Curtain installation in EVAR(Endovascular Aneurysm Repair) surgical operation using C-arm원문보기
In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation consider...
In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction.
In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction.
* AI 자동 식별 결과로 적합하지 않은 문장이 있을 수 있으니, 이용에 유의하시기 바랍니다.
문제 정의
본 연구는 장시간 C-arm을 이용하는 EVAR 시술 시, 수술 Staff가 받는 피폭선량에 대한 연구와 ,그 감소 방안의 연구로써, 수술 Table Side에 납 커튼을 설치하는 방법을 사용한 후 수술 Staff의 ESD(Entrance Surface Dose)를 측정하여 피폭선량 감소의 유·무와 효율성에 대해 연구해 보고자 한다.
제안 방법
1) 수술 Table에 Supine Position으로 Phantom을 위치시키고 Under tube로 Setting한 C-arm의 Center 를 Phantom의 L-4 level에 위치시켜 실제의 EVAR 시술과 동일한 조건으로 실험하였다.
2) C-arm의 X-ray Tube를 Under Positioning한 Under Tube system이 Over Tube system 보다 환자와 술자의 피폭선량의 양이 적다는 사실이 많은 논문과 연구에서 확인 되었고8~9, 본원 OR에서도 Run time 이 긴 혈관외과 및 영상의학과의 혈관 조영술과 중재적 시술시 Uuder tube system을 사용하여 환자 및 시술자의 의료피폭이 가능한 낮게 되도록 시행하고 있기 때문에 본 연구에서도 실제로 사용하고 있는 Under tube system으로 Positioning하여 실험하였다.
2) X-ray tube에서 Phantom간 거리는 55cm으로 하고, X-ray tube 위치로부터 수직 방향으로 50cm 지점과(이하 A point), 시술 중 수술 Staff의 행동반경을 고려하여 다시 이 지점으로부터 수술 table과 수평 방향으로 50cm 지점의(이하 B point) 두 위치에서 지면으로부터 50cm, 100cm, 150cm 높이의 위치에 유리선량계 소자를 부착하여 Lead curtain의 설치 전과 후의 선량을 각각 측정하였다.
3) EVAR 시술의 평균 Runtime은 1306.4sec지만 장시간 장비 가동시 Heating unit의 상승으로 장비의 손상을 초래할 수 있으므로 안정을 기하기 위하여 600sec 동안의 선량을 각각 1분씩 분할 조사 후 선량을 측정하였다.
성능/효과
1) 2012년 3월 2일부터 7월 11일까지 본원 OR에서 C-arm을 이용하여 EVAR 시술을 받은 환자를 조사해 본 결과 총 11명이었고, 가장 긴 Runtime이 2514sec, 가장 짧은 Runtime은 251.7sec 이었으며 평균 Runtime은 1306.4 sec 이었다.
100cm 높이에서 A point 와 B point의 측정된 선량은 설치 전의 경우 0.850mGy, 0.320mGy이었고 설치 후의 선량은 0.808mGy, 0.296mGy로써 다소 선량의 감소가 확인 되었지만 50cm 높이에서 측정된 선량의 감소에 비교하였을 때 선량 감소의 정도가 확연히 줄어든 결과를 확인할 수 있다.
150cm 높이에서 측정한 선량 값은 설치 전과 후가 변화가 거의 없는 것으로 봐서 Lead curtain 설치가 선량의 변화에 거의 영향을 주지 않는 것을 확인할 수 있다.
038mGy이었다. 두 지점에서 50cm 높이의 설치 전과 후의 선량이 각각 약 75%와 91% 이상 감소된 걸 확인할 수 있다.
본 연구는 시술시간이 장시간인 EVAR 시술 중 수술 Staff의 방사선에 대한 피폭선량을 알아보고, 피폭으로 부터 보호받지 못하는 Lower Extremity의 방사선 피폭을 감소시켜 보자는 취지에서 시행되었고, 실험결과 피폭선량 감소의 확실한 효과를 확인할 수 있었다.
실험 결과 50cm 높이에서 수술 Staff의 활동 반경 내의 두 지점에서 확연한 피폭선량의 감소를 확인 할 수 있었고 이는 추가로 방사선 보호구등을 착용하지 않고도 방사선 피폭으로부터 노출 되어 있었던 Lower Extremity를 보호할 수 있는 방안을 마련해 주었다.
후속연구
또한 과거 세대의 C-arm에 비하여 장비의 크기가 Compact화 되어 지고 영상획득에 필요한 선량 또한 과거에 비해 상당히 감소하고 있기 때문에 향후 수술 중 C-arm의 사용빈도가 더욱 증가 할 것이라 예상된다. 따라서 방사선 관계 종사자로서 시대에 흐름에 발맞추어 관련지식과 술기의 정확성을 습득하여 환자와 수술 참가인원의 피폭선량을 감소시키는 방법을 꾸준히 연구 하고 발전시켜 나아가야 한다.
본 연구의 경우 Lead curtain으로 기존에 만들어져 있는 임산부의 복부를 보호하는 제품(60cm × 53cm)을 사용하였지만 보다 적합화 된 모형으로 제작하여 사용 한다면 더욱 광범위한 부위의 피폭을 방어할 수 있을 것이고 기존의 방사선 방어구와 병용하여 사용한다면 수술 참가인원의 신체 거의 모든 부분의 방사선 피폭에 대한 방어가 이루어 질 수 있을 것이다.
질의응답
핵심어
질문
논문에서 추출한 답변
C-arm의 사용 장점은?
현재 수술 중 C-arm의 사용은 수술 기구의 정밀한 삽입, 수술시간의 단축, 정확한 정복술, 비침습적 상태 에서 환자의 Anatomical Structure를 확인할 수 있는 장점 뿐만 아니라 Subtraction 영상의 획득, Tomography 영상의 획득, Bolus Chasing 등의 최첨단 기능의 탑재로 인해서 정형외과 뿐만 아니라 신경외과 마취과 혈관외과 등 다양한 임상과에서 그 사용빈도가 갈수록 늘어가고 있는 추세이다.
의료 방사선에 대한 사회적 관심과 인식도가 높아진 이유는?
현대 의료기술의 발달로 진단적 검사 또는 치료학적으로 의료방사선에 대한 대중의 접근성이 매우 높아 졌고 실제로 20%인 인류의 인공 방사선 피폭 중 의료용 방사선에 의한 피폭이 약 15% 이상 차지하고 있을 만큼 과거에 비해서 그 비중이 높아졌다.1 그에 따라 의료 방사선에 대한 사회적 관심과 인식도 또한 과거에 비해 매우 높아졌고 시대의 흐름에 발맞춰 환자 및 방사선 관계종사자의 피폭선량을 줄일 수 있는 방안을 꾸준히 연구하고 발달시켜 나아가고 있는 추세이다.
C-arm을 이용한 수술 중 투시의 장점은?
현대의 의료분야에서 Radiological image는 진단적으로 가장 중요한 도구중의 하나이고 그중 C-arm을 이용한 수술 중 투시는 Real time으로 환자의 Anatomic structure를 확인할 수 있고 정확한 술기와 수술시간을 단축시키는 장점 등으로 인해 정형외과의 정복술을 비롯하여 신경외과, 비뇨기과, 마취과, 혈관 의과의 중재적 시술 등에 이르기까지 그 사용범위와 사용 빈도가 매년 높아지고 있고 본원에서도 매년 사용빈도가 약 10%씩 증가하고 있는 추세이다.3 그러나 수술중 투시는 수술과정에 따라서 환자에게 가해지는 피폭의 양이 일반 x-ray 촬영 1회시 받는 선량에 비교하여 작게는 수배에서 크게는 수십 배에 이르기도 하며 이때 발생하는 X-ray의 1차선과 산란선 또한 수술에 참가하는 인원에게 노출되어 잠재적으로 방사선 장해의 위험 요소가 될 수 있으므로 OR(Operation Room)에서의 C-arm 사용 증가에 따라 이에 비례적으로 환자와 수술참가인원의 피폭선량이 증가하게 되는 문제가 대두 되었다.